Intended Use
This drill guide facilitates the placement of the implants at the correct angulations which helps assure that the final position of the multi-unit abutment will be perpendicular to the patient’s jaw.
Characteristics
- Durable titanium construction
- Adjustable and foldable, allowing to be parallel to and follow the contour of the crestal bone.
- Sterilizable.
Step 1
Make a 2.0 pilot hole in order to place the guide pin.
Step 2
After placement of the guide pin into the drilled site, the angle and position of the guide band can be adjusted to be parallel to the crestal bone.
Step 3
The titanium band can be bent in order to follow the arch of the crestal bone.
Step 4
Three different angulations are indicated on the outside of the guide 0°, 17° and 30°. These angulations match the angulations of the different C-Tech multi-unit screw retained abutments.
- Step 5
Placement of the screw retained abutments.