Open Tray Impression Post Diameter 3

INTENDED USE
Open tray impression technique.

CHARACTERISTICS
– Simple
– Slender emergence profile accommodates space limitations
– Guide screw can be tightened either by hand or with the SCS screwdriver
– High precision impression components give an exact replica of the intraoral situation
– Clear-cut tactile response from the prosthetic connection verifies proper seating of components.

NOTE
Open tray impression procedure requires a custom-made
tray with perforations.

STEP 1
Place the impression post accurately into the implant and handtighten the guide screw.

STEP 2
Make perforations in the custom-made impression tray (light cured resin) according to the individual situation so that the positioning screw of the impression post sticks out.

STEP 3
Take the impression using an elastomeric impression material (polyvinyl siloxane or polyether rubber).

STEP 4
Reposition and fix the analog in the impression using the screw.